Baan’s Equation: Conductance-Based Pressure–Volume Loops

Baan’s equation forms the mathematical and physiological foundation of conductance-based pressure–volume (PV) loop technology. This method allows real-time, beat-to-beat measurement of left ventricular volume inside the intact, beating heart—making it one of the most powerful tools for studying cardiac mechanics, ventricular-arterial coupling, and contractility.

Traditional PV measurements relied on imaging or indicator dilution techniques, which lack the temporal resolution needed to capture instantaneous ventricular mechanics. Conductance technology solved this limitation by transforming intracavitary electrical signals into continuous volume measurements.


The Core Concept of Conductance Volume Measurement

The conductance catheter contains multiple electrodes arranged along its length. A high-frequency, low-amplitude current is applied between outer electrodes, while inner electrodes measure the resulting voltage. Because blood conducts electricity, the measured electrical conductance is proportional to the volume of blood inside the ventricle.

However, not all measured conductance comes from blood. The myocardium, surrounding tissues, and intracardiac structures also conduct electricity. Baan’s equation mathematically separates true blood volume conductance from parallel conductance generated by surrounding tissues.


The Baan Equation

The classical Baan equation is written as:

V(t) = (1/α) · ρ · L² · [G(t) − Gp]

Where:

  • V(t) = Instantaneous ventricular volume
  • α (alpha) = Calibration factor correcting for field non-uniformity
  • ρ (rho) = Blood resistivity
  • L = Distance between sensing electrodes
  • G(t) = Measured total conductance at time t
  • Gp = Parallel conductance (non-blood conductance)

This equation converts raw electrical conductance into physiologically meaningful ventricular volume.


Parallel Conductance (Gp): The Major Source of Error

Parallel conductance represents electrical current flowing through myocardium and surrounding tissues rather than blood. If not corrected, this leads to systematic overestimation of ventricular volume.

Clinically and experimentally, Gp is most commonly determined using the hypertonic saline method. A bolus of hypertonic saline transiently increases blood conductivity without affecting tissue conductivity. The shift in conductance allows separation of blood and tissue components.

Once Gp is known, it is subtracted from total conductance to obtain true blood conductance.


The Alpha (α) Calibration Factor

Even after correcting for parallel conductance, conductance-derived volume still requires scaling. The electric field inside the ventricle is non-uniform, and catheter position affects signal accuracy.

The α factor is determined by comparing conductance-derived stroke volume with an independent reference method such as:

  • Thermodilution
  • Echocardiographic stroke volume
  • Aortic flow probe

This step aligns conductance volume with true physiological volume and ensures accurate absolute PV loop scaling.


How Baan’s Equation Enables Real-Time PV Loops

Once calibrated, the conductance system continuously converts conductance into instantaneous volume. When synchronized with high-fidelity left ventricular pressure signals (usually from a micromanometer), the result is a real-time PV loop for every cardiac cycle.

This enables direct measurement of:

  • End-systolic pressure–volume relationship (ESPVR)
  • End-diastolic pressure–volume relationship (EDPVR)
  • Stroke work
  • Contractility indices (Ees, PRSW)
  • Ventricular-arterial coupling (Ea/Ees)

None of these parameters can be measured directly using echocardiography alone.


Strengths of Conductance-Based PV Analysis

The major advantage of Baan-based PV loops is temporal resolution. Volume and pressure are captured at millisecond scale, allowing precise analysis of:

  • Isovolumic contraction and relaxation
  • Load-independent contractility
  • Beat-to-beat responses to preload and afterload changes
  • Drug effects in real time
  • Acute valvular interventions and ventricular unloading devices

This makes the technique uniquely powerful in both cardiovascular research and advanced translational physiology studies.


Limitations and Sources of Error

Despite its strengths, conductance PV analysis is not without limitations:

  • Catheter position sensitivity – off-axis positioning distorts volume
  • Assumption of uniform electric field – violated in dilated or asymmetric ventricles
  • Calibration drift – α may change with geometry and loading
  • Invasiveness – limits routine clinical use

Clinical and Research Applications

Baan’s equation is the backbone of PV loop analysis in:

  • Experimental heart failure models
  • Valvular heart disease hemodynamic research
  • Mechanical circulatory support studies
  • Cardiac pharmacology
  • Ventricular-arterial coupling assessment
  • Acute interventional cardiology research

In structural heart disease, conductance PV loops uniquely reveal how regurgitant lesions, stenosis, and ventricular remodeling alter intrinsic myocardial performance beyond what standard pressure or volume measurements can show alone.


Key Takeaway

Baan’s equation transformed cardiac physiology by enabling true real-time ventricular volume measurement inside the beating heart. By correcting for parallel conductance and calibrating field geometry, it converts electrical conductance into accurate physiological volume—forming the basis of modern pressure–volume loop analysis.

Without Baan’s equation, the load-independent assessment of contractility, compliance, and ventricular-arterial interaction would not be possible at the level of precision used in contemporary cardiovascular research.